Chapter 5 465
One-Button Measurement Functions
Meas Setup (Spectrum Emissions Mask—SEM)
One-Button Measurement Functions
SEM:OFFS:LIST:STAT 1
SEM:OFFS:LIST:STAT?
5.22.4.3 Stop Freq
Specifies the outer limit (frequency furthest from the carrier) for both segments of the
specified offset pair. The lower range is limited to the setting of Start Freq.
Key Path: Meas Setup,Offset/Limits
State Saved: Saved in instrument state.
Factory Preset: Dependent upon Radio Standard and device selected, refer to the default
tables for “Offset/Limits” on page 462 for this value.
Default Terminator: Hz
Range: The Start Freq (for that offset) plus 10 Hz to 100 MHz
Remote Command:
[:SENSe]:SEMask:OFFSet[n]:LIST:FREQuency:STOP <freq> ... [<freq>] (up to five
values)
[:SENSe]:SEMask:OFFSet[n]:LIST:FREQuency:STOP?
Remote Command Notes: Comma separated list of up to 5 values. n = 1 for BTS and n = 2
for MS. Default is BTS.
Example: SEM:OFFS:LIST:FREQ:STOP 4 MHz
SEM:OFFS:LIST:FREQ:STOP?
5.22.4.4 Sweep Time
Specifies the sweep time for the currently selected offset, and toggles this function between
Auto and Manual for each offset.
Key Path: Meas Setup,Offset/Limits
State Saved: Saved in instrument state.
Factory Preset: Dependent upon Radio Standard and device selected, refer to the default
tables for “Offset/Limits” on page 462 for this value.
Default Terminator: s (seconds)
Range: 1 ms to 2 ks
Remote Command:
[:SENSe]:SEMask:OFFSet[n]:LIST:SWEeptime <time> ... [<time>] (up to five