412 Chapter 5
One-Button Measurement Functions
Meas Setup (Harmonic Distortion)
One-Button Measurement Functions
Remote Command:
[:SENSe]:HARMonics:SWEeptime:STATe OFF|ON|0|1
[:SENSe]:HARMonics:SWEeptime:STATe?
[:SENSe]:HARMonics:SWEeptime <time>
[:SENSe]:HARMonics:SWEeptime?
Remote Command Notes: ON is equal to automatic (Auto). OFF is equal to manual
(Man).
Example: HARM:SWE:STAT OFF
HARM:SWE 100 ms
5.16.5 Range Table (On/Off)
PressRangeTabletospecifywhethertherangetableistobeusedornot.WhenRange Table
is set to Off the harmonics are measured. When Range Table issettoOnthefirstactive
range is considered to be the fundamental and the subsequent active ranges are used as
the harmonics. At least two ranges must be active if Range Table is set to On.
Key Path: Meas Setup
Factory Preset: See Table 5-1 on page 414.
Range: Off|On
History: Added with firmware revision A.03.00
Remote Command:
[:SENSe]:HARMonics:RTABle:STATe OFF|ON|0|1
[:SENSe]:HARMonics:RTABle:STATe?
Example: HARM:RTAB:STAT OFF