Chapter 4 285
Instrument Functions: P - Z
Trace/View
Instrument Functions: P - Z

4.9.3 Clear Write

Erases any data previously stored in the selected trace and continuously displays signals
duringthesweepoftheanalyzer.
Key Path: Trace/View
Remote Command:
See “Trace/View” on page 283.
Example: TRAC:MODE WRIT

4.9.4 Max Hold

Maintains the maximum level for each trace point of the selected trace (1, 2 or 3), and
updates each trace point if a new maximum level is detected in successive sweeps.
NOTE Pressing Restart, changing the vertical scale (Amplitude,Scale Type,LogorLin)
or turning averaging on (BW/Avg,Average (On)) restarts the held trace.
Key Path: Trace/View
Remote Command:
See “Trace/View” on page 283.
Example: TRAC:MODE MAXH

4.9.5 Min Hold

Maintains the minimum level for each trace point of the selected trace (1, 2 or 3), and
updates each trace point if a new minimum level is detected in successive sweeps.
NOTE Pressing Restart, changing the vertical scale (Amplitude,Scale Type,LogorLin)
or turning averaging on (BW/Avg,Average (On)) restarts the held trace.
Key Path: Trace/View
History: Added with firmware revision A.02.00
Remote Command:
See “Trace/View” on page 283.
Example: TRAC:MODE MINH