Chapter 5 353
One-Button Measurement Functions
Meas Setup (Adjacent Channel Power—ACP)
One-Button Measurement Functions
5.6.2 Avg Mode
Press Avg Mode to select the type of termination control used for the averaging function as
either Exp or Repeat. This determines the averaging action after the specified number of
measurements (average count) is reached.
•EXP (Exponential Averaging mode)—When you set Avg Mode to Exp, each successive data
acquisition after the average count is reached is exponentially weighted and combined
with the existing average. Exponential averaging weights new data more than old data,
which facilitates tracking of slow-changing signals. The average will be displayed at the
end of each sweep.
•Repeat—When you set Avg Mode to Repeat, after reaching the average count, all previous
result data is cleared and the average count is set back to 1.
Key Path: Meas Setup
State Saved: Saved in instrument state.
Factory Preset: EXPonential
History: Added with firmware revision A.02.00.
Remote Command:
[:SENSe]:ACPower:AVERage:TCONrol EXPonential|REPeat
[:SENSe]:ACPower:AVERage:TCONrol?
Example: ACP:AVG:TCON EXP
ACP:AVG:TCON?
5.6.3 Chan Integ BW
Press Chan Integ BW to specify the range of integration used in calculating the power in the
main channel.
Key Path: Meas Setup
Key Notes: If RRC Filter is on, the actual integration bandwidth used is the displayed
integration bandwidth multiplied by (1 + Filter Alpha).
State Saved: Saved in instrument state.
Factory Preset: 2.0 MHz or as defined by the selected radio standard.
Default Terminator: Hz
Knob Increment: Span/50 if non-zero span.
RBW/100 if zero span.
Step Key
Increment: RBW if CF Step is set to Auto with zero span.