One-Button Measurement Functions
Restart
One-Button Measurement Functions

5.4 Restart

This function restarts a previously paused measurement at the beginning. If the
measurement is active, it will stop it as soon as possible and restart it from the
beginning.If no measurement is active and Sweep (Single) is selected, a new sweep is
initiated.
When in Average (On) mode, (BW/Avg,Average) the averaging function is restarted (the trace
is reset and the average number is reset to zero).
Key Path: Front-panel key. It can also be found under Meas Control.
Remote Command:
:INITiate:RESTart
Remote Command Notes: This command is equivalent to sending an :ABORt command
followed by an :INITiate[:IMMediate] command. See “Abort the Sweep or
Measurement (Remote Command Only)” on page 331. for more information.