Chapter 6 541
Programming Fundamentals
Using the LAN to Control the Instrument
Programming Fundamentals
are set correctly.
Cannot access the file system via ftp
If you get a "connection refused" message, try the following
solutions:
If the power to the instrument was just turned on, make sure that
you wait about 25 seconds before attempting the connection.
If you get a "connection timed out" message
Verify the LAN connection between your computer and the
instrument. Refer to "If you cannot connect to the instrument"
earlier in this section.
Cannot telnet to the command parser port
For a "connection refused" message
Check the telnet port number from the front panel keys.
For a "connection timed out" or "no response from host" message
Verify the LAN connection between your computer and the
instrument. Refer to "If you cannot connect to the instrument"
earlier in this section.
For a "connection refused" or "no response from host" message
If the instrument was just turned on, make sure that you wait
about 25 seconds before attempting the connection.
An "operation timed-out" message
Check the LAN connection between the computer and the
instrument. Refer to "If you cannot connect to the instrument" in
this section.
Increase the file time-out value on your PC or workstation.
Cannot access internal web pages or import graphic images
when using a point-to-point connection
Disable the use of proxy servers. You may have to specify this in a
number of locations, depending on the operating system and
software you are using.
Disable the use of cached copies of web pages to ensure that you
always get a new copy of the instrument’s screen image.
If all else fails
Contact your network administrator.
If you still cannot solve the problem, contact an Agilent Service
Center for repair information.