456 Chapter 5
One-Button Measurement Functions
Meas Setup (Occupied Bandwidth—OBW)
One-Button Measurement Functions
Remote Command:
[:SENSe]:OBWidth:XDB <dB value>
[:SENSe]:OBWidth:XDB?
Example: OBW:XDB 50 dB
OBW:XDB?

5.21.7 Optimize Ref Level

Sets the input attenuator to optimize the robustness of the measurement, which is its
freedom from errors due to input compression.
NOTE There will always be a minimum of 6 dB of attenuation set to protect the
analyzer input.
Key Path: Meas Setup
State Saved: Saved in instrument state.
Factory Preset: n/a
Remote Command:
[:SENSe]:POWer[:RF]:RANGe:AUTO ONCE
Example: POW:RANG:AUTO ONCE