Chapter 2 69
Instrument Functions: A - L
AMPLITUDE / Y Scale
Instrument Functions: A - L
Sets the Attenuation to 10 dB. The only valid suffix is dB. If no suffix is
sent, dB is assumed.
If a value >5 is entered, 10 is used.
If a value 5 is entered, 2 is used for PSA, or 5 is used for ESA.

2.0.13 Max Mixer Lvl

Enables you to set the relationship between the highest signal that can be displayed (the
reference level) and the input attenuation. The relationship applies whenever the
Attenuation is set to Auto. The relationship is that the attenuation is given by reference
level minus the max mixer level. For example, as the reference level changes, the input
attenuator changes to ensure that a signal at the reference level does not exceed the Max
Mixer Lvl setting.
Key Path: AMPLITUDE / Y Scale
State Saved: Saved in Instrument State
Factory Preset: –10 dBm
Range: –50 dBm to –10 dBm
Remote Command:
[:SENSe]:POWer[:RF]:MIXer:RANGe[:UPPer] <power>
[:SENSe]:POWer[:RF]:MIXer:RANGe[:UPPer]?
Example: POW:MIX:RANG -15 dBm