Chapter 5 349
One-Button Measurement Functions
Single
One-Button Measurement Functions

5.5 Single

If the analyzer is in continuous sweep mode and not in a measurement (Measure,Meas Off),
pressing Single changes the sweep control to single sweep, and executes a sweep after the
trigger condition is met. If the analyzer is already in single sweep, pressing Single executes
a new sweep after the trigger condition is met.
Some instrument settings require more than one sweep to complete the measurement (see
BW/Avg,Average), or if you have selected a measurement from the functions under the
MEASURE key, this function sets the trigger system to be initiated only once. In this case the
trigger condition can be met only once and then all the necessary sweeps will be executed
to make the measurement or complete the averaging function.
With Average on (BW/Avg,Average (On)), pressing Single resets the average trace and starts
the average again from a count of zero. Sweeps are averaged until N sweeps are then taken
(where N is the average number), and then the sweep is halted.
State Saved: Single sweep setting is saved in instrument state
Factory Preset: Continuous
Remote Command:
:INITiate[:IMMediate]
*TRG
Remote Command Notes: Use the :TRIGger[:SEQuence]:SOURce command to select the
trigger source.
See also the Single/Continuous sweep function in “Sweep” on page 242 and
the command INITiate:CONTinuous ON|OFF.
See also the *TRG command information in the STATus section.
Example: *TRG
TRIG:IMM