Chapter 5 445
One-Button Measurement Functions
Meas Setup (Multi-Carrier Power—MCP)
One-Button Measurement Functions
Remote Command:
[:SENSe]:MCPower:FILTer[:RRC]:ALPHA <number>
[:SENSe]:MCPower:FILTer[:RRC]:ALPHA?
Example: MCP:FILT:ALPHA .33
MCP:FILT:ALPHA?
5.19.12 Noise Correction
Pressing Noise Correction turns noise correction on or off.WhenyousetNoise Correction to On,
a calibration of the noise floor is performed and used to correct for analyzer noise floor
contribution to measurement levels, increasing dynamic range.
Key Path: Meas Setup,More
Key Notes: This key is not available when Meas Setup,Method (RBW) is selected or when
signal tracking is on.
State Saved: Saved in instrument state.
Factory Preset: Off
History: Added with firmware revision A.02.00.
Remote Command:
[:SENSe]:MCPower:CORRection:NOISe[:AUTO] OFF|ON|0|1
[:SENSe]:MCPower:CORRection:NOISe[:AUTO]?
Remote Command Notes: The noise correction feature is not available when the
measurement method is RBW (Meas Setup,Method (RBW) or when signal
tracking is on.
Example: MCP:CORR:NOIS 1
MCP:CORR:NOIS?