Chapter 5 499
One-Button Measurement Functions
Meas Setup (Spurious Emissions)
One-Button Measurement Functions
Increment: 10
Range: –180 dBm to 20 dBm
Remote Command:
[:SENSe]:SPURious:POWer[:RF]:LEVel <integer>
[:SENSe]:SPURious:POWer[:RF]:LEVel?
[:SENSe]:SPURious:POWer[:RF]:RANGe:AUTO OFF|ON|0|1
[:SENSe]:SPURious:POWer[:RF]:RANGe:AUTO?
Example: SPUR:POW:RF:LEV 10
SPUR:POW:RF:RANG:AUTO ON

5.27.7 Fast Spurious Meas

Pressing Fast Spurious Meas turns the fast spurious measurement test on or off. Pressing Fast
Spurious Meas to On provides a faster method of execution as with fast spurious testing spurs
above the limit line are reported. Any spurs reported outside the limit will cause the
measurement to fail. See β€œAbs Start Limit” on page 493 for more information.
Key Path: Meas Setup
Saved State: Ref Level state is saved to instrument state. Ref Level value is
recalculated for each active range the value is not stored in instrument
state.
Factory Preset: Off
Range: Off|On
History: Add with firmware revision A.03.00
Remote Command:
[:SENSe]:SPURious:FSMeas OFF|ON|0|1
[:SENSe]:SPURious:FSMeas?
Example: SPUR:FSM ON