Chapter 5 497

One-Button Measurement Functions

Meas Setup (Spurious Emissions)

One-Button Measurement Functions

The query for this parameter will always return 20 values.

[:SENSe]:SPURious[:RANGe][:LIST]:PEAK:THReshold?

5.27.4 Meas Type

Specifies the measurement type (Examine or Full). This parameter is coupled to the average

mode. If you select Examine, the measurement sets Avg Mode to Exp. If you select Full, the

measurement sets Avg Mode to Rep. The behavior of each measurement type is described

below.

Key Path: Meas Setup

State Saved: Saved in instrument state.

Factory Preset: Examine

Range: Examine|Full

Single Continuous
No Spurs Spurs No Spurs Spurs
Examine All active ranges are
measured.
All active ranges are
measured and the found
spurs are reported.
All active ranges are
measured.
All active ranges are
measured and the found
spurs are reported.
Upon completion, the
measurement is set to
the idle state.
Upon completion, the
measurement is set to
the idle state.
Upon completion, the
analyzer remains set to
the last range checked
with an active trace.
Upon completion the
analyzer is set to the
range containing the
worst spur found and
continually sweeps this
range.
The message No Spurs
is displayed.
The trace containing the
worst spur is restored.
The message No Spurs
is displayed.
Spur is enabled (no
longer greyed out).
Spur is enabled (no
longer greyed out).
A marker is added, set
to the frequency of the
worst spur.
A marker added, set to
the frequency of the
worst spur.
Full All active ranges are
measured.
All active ranges are
measured and the found
spurs are reported.
Measurement
continually cycles
through all active
ranges.
All active ranges are
measured and the found
spurs are reported.
Upon completion, the
measurement is set to
the idle state.
Upon completion, the
measurement is set to
the idle state.
On each cycle of the
active ranges, the report
on spurs found is reset.
This ensures that
remote queries retrieve
the trace data that
matches the currently
displayed results.
The message No Spurs
is displayed.
The trace of the last
active range is
displayed.