MODEL | SECTION 2.0 |
| INSTALLATION |
2.3.4 Inductive Loops.
The Model
2.3.5 Sensor Selection.
All Rosemount Analytical contacting conductivity sensors with PT100 RTD or PT1000 RTD are compatible with the Model
Choose an inductive conductivity sensor that is appropriate for your process conditions and range of conductivity meas- urement.
TABLE 2-1. Model 5081-T Sensor Selection
RECOMMENDED RANGES FOR TOROIDAL SENSORS
| Conductivity Sensor |
|
|
|
|
|
|
| Model Number | 226 | 228 | 225 | 222 (1in.) | 222 (2 in.) | 242 |
|
|
|
|
|
|
|
|
| Nominal Cell Constant | 1.0 | 3.0 | 3.0 | 6.0 | 4.0 | * |
| Min. Conductivity (∝S/cm) | 50 | 200 | 200 | 500 | 500 | 100* |
|
|
|
|
|
|
|
|
| Max. Conductivity (∝S/cm) | 1,000,000 | 2,000,000 | 2,000,000 | 2,000,000 | 2,000,000 | 2,000,000* |
|
|
|
|
|
|
|
|
* Model 242 values depend on sensor configuration and wiring.
NOTE: Values shown are for 25°C conductivity with a temperature slope of 2% per degree C. The maximum range value will be lower for solutions with a higher temperature slope. Minimum conductivity depends on sensor.
RECOMMENDED SENSORS:
Model 222
Model 225
Model 226 Submersion/
Insertion
Model 228 Submersion/
Insertion/
Retractable
Model 242
*Model
11