MODEL | SECTION 7.0 |
| PROGRAMMING |
7.7 RANGE
7.7.1 Purpose
This section provides the steps to select automatic ranging or a specific fixed range of measurement. Five specific con- ductance ranges are selectable. Setting the Model 5081 to a fixed range reduces response time.
The following conductance ranges are available:
Setting | Measurement Range | Over Range Warning Limit |
|
|
|
Auto | 0uS to 1400mS | None |
|
|
|
1400mS | 550mS to 1400mS | None |
|
|
|
550mS | 200mS to 550mS | 570mS |
|
|
|
200mS | 33mS to 200mS | 207mS |
|
|
|
33mS | 3000uS to 33mS | 34mS |
|
|
|
3mS | 0uS to 3000uS | 3400uS |
|
|
|
NOTE: The selection between automatic ranging or a specified fixed range of measurement range can ONLY be done using the IRC/Infrared Remote Controller. This selection cannot be done via HART or FOUNDATION Fieldbus hosts or configurators.
7.7.2 Procedure
1.Press PROG.
2.Press NEXT five times.
3.Press ENTER. The default setting Auto will appear. This indicates that Model 5081 is in auto ranging mode.
4.To set a fixed conductance range, press the down arrow key until the desired measurement range appears. 1400mS, 550mS, 200mS, 33mS and 3mS will appear when pressing the down arrow successively.
5.When the desired range is reached, press ENTER. This disables auto ranging and limits the measurement to the selected range.
6.Press NEXT to move to the next Programming menu item.
7.8 DEFAULT
7.8.1 Purpose
This section describes how to erase ALL
7.8.2 Procedure
PROGRAM
G12->[U
EXIT NEXT ENTER
PROGRAM
2-FUQSA P9
EXITENTER
1.Press PROG on the remote controller.
2.Press NEXT until the dEFAULt appears in the display. Press ENTER.
3.Use or to toggle between nO and yES. With yES showing, press ENTER to return to factory default settings.
57