MODEL 5081-T

SECTION 9.0

 

OPERATION WITH MODEL 375

9.3Operation

9.3.1Off-line and On-line Operation

The Model 375 Communicator features off-line and on-line communications. On-line means the communicator is connected to the transmitter in the usual fashion. While the communicator is on line, the operator can view meas- urement data, change program settings, and read diagnostic messages. Off-line means the communicator is not connected to the transmitter. When the communicator is off line, the operator can still program settings into the communicator. Later, after the communicator has been connected to a transmitter, the operator can transfer the programmed settings to the transmitter. Off-line operation permits settings common to several transmitters to be easily stored in all of them.

9.3.2Making HART related settings from the keypad

Calibrate

Hold

Program

Display

 

 

 

 

Output

Temp

Measurement

>>

 

 

 

 

Security

HART

 

>>

 

 

 

 

DevID

PollAddrs

Burst

Preamble

 

 

1.Press MENU. The main menu screen appears. Choose Program.

2.Choose >>.

3.Choose HART.

4.To display the device ID, choose DevID. To change the polling address, choose PollAddrs. To make burst mode settings, choose Burst. To change the preamble count, choose Preamble.

9.3.3Menu Tree

The menu tree for the Model 375 HART communicator is on the following page. The menu tree for the Model 375 FOUNDATION Fieldbus communicator immediately follows.

Note: Model 375 Communicator fully supports FOUNDATION Fieldbus on Model 5081-T-FF.