Emerson 5081-T instruction manual RTD Resistance Values, Formula, Example

Models: 5081-T

1 98
Download 98 pages 25.35 Kb
Page 87
Image 87

MODEL 5081-T

10.6 RTD RESISTANCE VALUES

Table 10-4 is a ready reference of RTD resistance values at various temperatures. These are used for test and evaluation of the sensor.

NOTE

Resistance values are read across the RTD element and are based on the manu- facturer’s stated values (±1%). Allow enough time for the RTD element in the sensor to stabilize to the surrounding temperature (10 min).

SECTION 10.0

DIAGNOSIS AND TROUBLESHOOTING

Table 10-4. RTD Resistance Values.

Temperature

Pt-100

Pt-1000

(°C)

Resistance (ohms)

Resistance (ohms)

0

100.00

1000

10

103.90

1039

20

107.79

1078

25

109.62

1096

30

111.67

1117

40

115.54

1155

50

119.40

1194

60

123.24

1232

70

127.07

1271

80

130.89

1309

90

134.70

1347

100

138.50

1385

Use the following formula to determine the appropriate resistance value to use to simulate a conductivity value:

FORMULA:

cell constant value x 1,000,000

= resistance in ohms

desired simulated conductivity in s/cm

 

EXAMPLE:

.01 x 1,000,000

10 s/cm

= use 1,000 ohm resistance

FIGURE 10-5. Conductivity Determination

81

Page 87
Image 87
Emerson 5081-T instruction manual RTD Resistance Values, Formula, Example