Emerson 5081-T instruction manual Display, G5-A, Hcl, H2SO4L, H2SO4H

Models: 5081-T

1 98
Download 98 pages 25.35 Kb
Page 60
Image 60

MODEL 5081-T

SECTION 7.0

 

PROGRAMMING

7.4 DISPLAY

7.4.1 Purpose

This section describes how to do the following:

1.Configure the transmitter to measure conductivity, resistivity, or set up a custom curve

2.Set the temperature units to °C or °F

3.Set the output to current or percent of full scale

4.Enter a security code.

7.4.2 Definitions

1.MEASUREMENT. The transmitter can be configured to measure conductivity in mS/cm or resistivity in MegOhms, or configured with a 3-5 point custom curve for special applications.

2.OUTPUT CURRENT. The transmitter generates a 4-20 mA output signal directly proportional to the conductivity or resis- tivity of the sample. The output signal can be displayed as current (in mA) or as percent of full scale.

5. SECURITY CODE. The security code unlocks the transmitter and allows access to all menus.

7.4.3 Procedure

PROGRAM

G5<:[-A

EXIT NEXT ENTER

PROGRAM

UA:1 /QPGVF

EXITENTER

1.Press PROG on the remote controller.

2.Press NEXT until the diSPLAy submenu appears. Press ENTER.

3.Press ￿ or ￿ to display the desired measurement. Press ENTER to save.

 

COnduc

Conductivity

 

 

nAOH

Sodium Hydroxide 0-15%

 

 

HCL

Hydrochloric Acid 0-16%

 

 

H2SO4L

Sulfuric Acid 0-30%

 

 

H2SO4H

Sulfuric Acid 96-99.7%

 

 

CuSt

Custom Curve

 

 

 

 

 

4.tEMP will appear. Press ￿ or ￿ to display the desired temperature reading, C or F. Press ENTER.

5.OUtPUt will appear. Press ￿ or ￿ to display the desired 4-20 output value, Cur or %. Press ENTER.

6.COdE will appear. Using the arrow keys on the remote control, enter the desired 3-digit security code for accessing the Calibration, Program, and Diagnostic functions via the Remote Control handheld. Press ENTER. The dISPLAY mnemonic will appear.

54

Page 60
Image 60
Emerson 5081-T instruction manual Display, G5-A, Hcl, H2SO4L, H2SO4H