Emerson Electrical Installation, Wiring Model 242 sensor to Model 5081-T transmitter

Models: 5081-T

1 98
Download 98 pages 25.35 Kb
Page 22
Image 22

MODEL 5081-T

SECTION 3.0

 

WIRING

3.2 ELECTRICAL INSTALLATION

All Rosemount Analytical contacting conductivity sensors with PT100 RTD or PT1000 RTD are compatible with the Model 5081-T transmitter. Please refer to Figures 3-5 thru 3-7 for appropriate sensor to transmitter wiring. The sensor cable should be routed through the left inlet closest to the connector.

NOTE

Optimum EMI/RFI immunity may be achieved on sensors whose interconnecting cable has an outer braided shield by utilizing a cable gland fitting that provides for continuity between the braided shield and the transmitter enclosure. An equivalent conduit connector may also be used if the sensor cable is to be enclosed in conduit.

FIGURE 3-5. Wiring Model 242 sensor to Model 5081-T transmitter

16

Page 22
Image 22
Emerson instruction manual Electrical Installation, Wiring Model 242 sensor to Model 5081-T transmitter