Emerson Process Management GmbH & Co. OHG6-8
X-STREAM XE
Instruction Manual
HASXEE-IM-HS
06/2010
6.2.2.3 Advanced Calibration Menu
6.2.2 Control Menu
Note!
This menu is only available if “Valves” in IN-
STALLED OPTIONS is set to a value other
than none.
Note!
This menu is also available in single-channel
units.
In this case, the 3rd and 4th lines will start a
zero or span calibration, while the 5th line
starts the same procedure as the 2nd.
Start and control valve supported calibration
procedures
Control..
Advanced calibration..
If the system is setup accordingly, the access
code for level 1 must be entered to gain ac-
cess to this menu.
Advanced Calibration 1of2
Cancel!
Zero&Span Single..
Zero All!
Span All!
Zero&Span All!
Programmed Sequence!
Blowback!
Press enter in this line to cancel any ongoing
calibration
Highlight any of the next 3 lines and press
enter to start the related calibration proce-
dure:
- zero all channels
- span all channels
- zero and span all channels
Press enter in this line to start a blowback pro-
cedure for all channels
Press enter in this line to start a programmed
calibration sequence;
6-39 for information
on how to program a sequence.
Zero & span a single channel 6-5