Emerson Process Management GmbH & Co. OHG 3-7
X-STREAM XE
Instruction Manual
HASXEE-IM-HS
04/2010
3Principles
3.2 Oxygen MeasurementTab. 3-2: Solvent Resistant Paramagnetic Sensor -
Approved Solvents
Component
Measuring cell type
Solvent resistant Corrosion resistant
(Chlorine, dry)
Case SS 1.4572 SS 1.4573
Pole nucleus Tantalum
Mirror Glass, Rhodium
Tension band Platinum alloy
Loop wire Platinum alloy
Supporting
wire Platinum alloy
Cylinder Glass
Cylinder
bushing Ceramics
Dumbbell Glass
Taring Epoxy Epoxy
Compound
material Plumb bob, Epoxy Epoxy
Seals Kalrez Kalrez
Tab. 3-3: Solvent Resistant Paramagnetic Sensor -
Medium Affected Materials
3.2.1.2 Applications With Corrosive Or Solvent ComponentsSpecial paramagnetic oxygen sensors are available to measure gases, containing cor-rosive or solvent components. See below tables for further information on approved solvents, and medium affected materials.The table below shows, how accompanying gases interfere the paramagnetical oxygen measurement. If the concentration of such gases is already given at time of enquiry, this interference may be taken into account during factory startup and thus minimized (option).Tab. 3-4: Solvent Resistant Paramagnetic Sensor -
Cross Interference by Accompanying
Gases
100 % Gas Zero-level effect % O2
Argon Ar -0.22
Carbon Dioxide CO2-0.27
Carbon Monoxide CO +0.01
Ethane C2H6-0.46
Ethene C2H4-0.26
Helium He +0.3
Hydrogen H2+0.24
Methane CH4-0.2
Neon Ne +0.13
Nitrogen N2+0.00
Nitrogen Dioxide NO2+28.0
Nitrogen Oxide NO +43.0
Nitrous Oxide N2O -0.2
Propane C3H8-0.86
Propene C3H6-0.55
Approved solvents
(inclusive accompanying disturbing
components)
Component Concen-
tration
Component Concen-
tration
Acetic acid 0.1- 20 %
Acetone 0.1- 20 %
Acrolein 0.1- 20 %
Aromatics 0.1- 20 %
Butadiene 0.1- 20 %
Butadiene-1 0.1- 20 %
Butadiene-2 0.1- 20 %
C2H2 0.1- 20 %
C4H8 0.1- 20 %
C5 0.1- 20 %
C6H12 0.1- 20 %
CH3COOH 0.1- 20 %
Cyclohexane 0.1- 20 %
Cyclohexanon 0.1- 20 %
Dimethyl sulde 0.1- 20 %
Ethanol 0.1- 20 %
Ethene 0.1- 20 %
Ethylene 0.1- 20 %
Ethylene oxid 0.1- 20 %
Heptane 0.1- 20 %
Hexane 0.1- 20 %
Isopropanol 0.1- 20 %
Methanol 0.1- 20 %
Methyl acetate 0.1- 20 %
Methylethylketone 0.1- 20 %
Methylmercaptane 0.1- 20 %
Propadiene 0.1- 20 %
Propene 0.1- 20 %
Propylen oxide 0.1- 20 %
Propylene 0.1- 20 %
Toluene 0.1- 20 %
Vinyl acetate 0.1- 20 %
Vinyl acetylene 0.1- 20 %
Xylene 0.1- 20 %
i-Butyr acid 0.1- 20 %
i-Butyr aldehyd 0.1- 20 %
i-Propylformiat 0.1- 20 %
n-Butane 0.1- 20 %
Conditions
· Single or summarized concentrations do not
exceed 20 %
· Gas passes gas cooler prior to entering the
analyzer
· Gas dew point at max. 5 °C
Solvent resistant sensors are consumables!