3Principles
3.6 Measurement Specication
3.4 Measurement SpecicationSample gas components and measuring ranges (standard congurations) Note!The following table gives an overview only. Not all data is applicable to all analyzer variations. The sample gas(es) and measuring ranges for your specic analyzer are given by the order acknowledgement and on the analyzer‘s name plate label.

Tab. 3-7: Gas Components and Measuring Ranges, Examples

1 More than 60 gases are detectable; other
components and congurations on request
2 Dew point below ambient temperature
3 Higher concentrations decrease sensor lifetime
4 Non-standard specications for lowest range
5 Special "renery" application with 0 - 1% H2 in N2 available
Gas component 1Lowest measuring range Highest measuring range
Acetone CH3COCH30 - 200 ppm 40 - 3 %
Acetylene C2H20 - 3 % 0 - 100 %
Ammonia NH30 - 100 ppm 0 - 100 %
Argon Ar 0 - 50 % 0 - 100 %
Carbon dioxide CO20 - 5 ppm 40 - 100 %
Carbon monoxide CO 0 - 10 ppm 4 0 - 100 %
Chlorine Cl20 - 300 ppm 0 - 100 %
Ethane C2H60 - 1000 ppm 0 - 100 %
Ethanol C2H5OH 0 - 1000 ppm 0 - 10 %
Ethylene C2H40 - 400 ppm 0 - 100 %
Helium He 0 - 10 % 0 - 100 %
Hexane C6H14 0 - 300 ppm 0 - 10 %
Hydrogen H2
0 - 2 %
5
0 - 100 %
Methane CH40 - 100 ppm 0 - 100 %
Methanol CH3OH 0 - 1000 ppm 0 - 10 %
n - Butane C4H10 0 - 800 ppm 0 - 100 %
Nitrogen dioxide NO20 - 25 ppm 0 - 10 %
Nitrogen monoxide NO 0 - 100 ppm 0 - 100 %
Nitrous oxide N2O 0 - 100 ppm 0 - 100 %
Oxygen (electrochemical) O20 - 5 % 0 - 25 % 3
Oxygen (paramagnetic) O20 - 1 % 4 0 - 100 %
Propane C3H80 - 1,000 ppm 0 - 100 %
Propylene C2H60 - 10 % 0 - 100 %
Sulfur dioxide SO20 - 25 ppm 0 - 100 %
Sulfur hexauoride SF60 - 20 ppm 0 - 2 %
Toluene C7H80 - 300 ppm 0 - 5 %
Vinyl chloride C2H3Cl 0 - 2 % 0 - 2 %
Water vapor 2H2O 0 - 1000 ppm 0 - 8 %