Emerson Process Management GmbH & Co. OHG A-39

X-STREAM XE
Instruction Manual
HASXEE-IM-HS
04/2010
AAppendix

PLCQuickReferenceCard

Rev.201004
ChannelActionsPoolIndex
Usageexample:CALL A103; Start span cal for channel 1, if register changed to TRUE
#Replace ‘x’ in table by channel # (1 … 5)
ANR Description  ANRDescription
x01Nonex12MaintRequ
x02ZeroCalXx13FctCheck
x03SpanCalx14SampleGas
x04ZeroSpanCalx15ZeroGas
x05Cancelx16SpanGas1
x06Range1x17SpanGas2
x07Range2x18SpanGas3
x08Range3x19SpanGas4
x09Range4x20AllClosed
x10Failurex21Blowback
x11OffSpec
ChannelDigitalOutputPoolIndex
Usageexample:LOAD S332; Load press low alarm state of channel 3 into register
#Replace ‘x’ in table by channel # (1 … 5)
SNr Description  SNRDescription
x01Off x21Conc.Hi
x02On x22Conc.HiHi
x03Heartbeat  x23AverageLoLo
x04Failure  x24AverageLo
x05OffSpec  x25AverageHi
x06MaintRequ  x26AverageHiHi
x07FctCheck  x27TemperatureLoLo
x08Calibrating  x28TemperatureLo
x09Zeroing  x29TemperatureHi
x10Spanning  x30TemperatureHiHi
x11ZeroFailed  x31PressureLoLo
x12SpanFailed  x32PressureLo
x13RangeUnderflow  x33PressureHi
x14RangeOverflow  x34PressureHiHi
x15Range1  x35FlowLoLo
x16Range2  x36FlowLo
x17Range3  x37FlowHi
x18Range4  x38FlowHiHi
x19Conc.LoLo  x39Off
x20Conc.Lo  x40Off
A.7 PLC Quick Reference