Emerson Process Management GmbH & Co. OHG6-40
X-STREAM XE
Instruction Manual
HASXEE-IM-HS
06/2010
6.2.3.2.6 Setup Calibration Interval Time
6.2.3 Setup Menu
Select the procedure(s) you want to congure
to be carried out on a regular (interval time)
basis.
Note!
All 4 lines in this menu link to submenus of
a similiar design, exemplied in the following
section.
Setup..
Calibration..
Interval Times..
Interval Times
Zero All..
Zero&Span All..
Programmed Sequence..
Blowback All..
Note!
For information about how to access menus
exemplied in this section, above.
...
Enable: Disabled
Interval: 24 h
Start Time..
Next: --
Time 6/22/09 15:33
6.2.3.2.6.1 Setup an Interval Time
Depending on the procedure selected on the
previous menu page, the title shows 'Zero All',
'Zero&Span All', 'Programmed Sequence' or
'Blowback All'.
Enable or Disable interval times for the se-
lected procedure
Specify the interval between two proce-
dures.
Accepted values: 1 ... 10,000 h
Shows the time for the next start of procedure,
based on the current settings.
Current time.
Note!
This line is available only if “Valves” in IN-
STALLED OPTIONS is set to a value other
than none (
6-88 )
.