Emerson Process Management GmbH & Co. OHG 3-15
X-STREAM XE
Instruction Manual
HASXEE-IM-HS
04/2010
3Principles
3.6 Measurement Specication
Special Performance Specications for Gas Purity Measurements (ULCO & ULCO2)

Tab. 3-9: Special Performance Specications for Gas Purity Measurements

All these performance data is veried during the manufacturing process for each unit by the following tests: Linearization and sensitivity test Long term drift stability test Climate chamber test Cross interference test (if applicable)
A61: 0 - 10 ... 100 ppm CO
A62: 0 - 20 ... 100 ppm CO
B62: 0 - 10 ... 100 ppm CO2
B61: 0 - 5 ... 100 ppm CO2
Noise < 1 % 1 2 < 1 % 1 2 resp. < 0.1 ppm 1 2 9
Detection limit < 2 % 1 2 < 2 % 1 2 resp. < 0.2 ppm 1 2 9
Linearity < +1 % 1 2 < +1 % 1 2
Zero-point drift < +2 % 1 2 3 < +2 % 1 2 3 resp. < 0.2 ppm 1 2 3 9
Span (sensitivity) drift < +2 % 1 2 4 < +2 % 1 2 4 resp. < 0.2 ppm 1 2 4 9
Repeatability < +2 % 1 2 < +2 % 1 2 resp. < 0.2 ppm 1 2 9
Response time (t90) < 10 s 7< 10 s 7
Permissible gas ow 0.2 - 1.5 l/min. 0.2 - 1.5 l/min.
Inuence of gas ow < +2% 1 2 < +2% 1 2
Maximum gas pressure 10 < 1,500 hPa abs. (< 7 psig) < 1,500 hPa abs. (< 7 psig)
Inuence of pressure
– At constant temperature < 0.1 % per hPa 5< 0.1 % per hPa 5
– With pressure compensation 8< 0.01 % per hPa 5< 0.01 % per hPa 5
Permissible ambient temperature 5 to +40 °C (41 to 104 °F) 5 to +40 °C (41 to 104 °F)
Inuence of temperature
(at constant pressure)
– On zero point < +2 % per 10 K 6< +2 % per 10 K 6 resp. < 0.2 ppm per 10 K 6 9
– On span (sensitivity) < +2 % per 10 K 6< +2 % per 10 K 6 resp. < 0.2 ppm per 10 K 6 9
Thermostat control None None
1Related to full scale
2Constant pressure and temperature
3Within 24 h; daily zero calibration requested
4Within 24 h; daily span calibration recommended
5Related to measuring value
6Temperature must not change > 10 K within 1 h
7From gas analyzer inlet at gas ow of 1.0 l/min
8 Barometric pressure sensor is required
9Applies to measuring range < 10 ppm
10 Limited to atmospheric if internal sample pump