Adjust Conversion Input1
Mode: 4..20mA
ADC Output 8388608
Input Normalized to 0...1
Coeff a0: 1
Coeff a1: 2
Coeff a2: 3
Calc. Input Value 1013
Status Good
Select the type of signal, provided by the
external source.
Available options:
0..10V, 0..1V, 4..20mA: voltage or current
signal is applied
Off: Input is switched off
Test: the instrument internally generates a
step-like signal for testing purposes.
Enter the coefcients for your input signal
here, considering
- it is for a cubic equation max.
- it is normalized to 0...1 (see note in menu)
Shows the internal ADC output (informative)
Shows the calculated input value, based on
the coefcients a0..a2, for comparison with
the real input value. If both do not match,
correct the coefcients.
Input signal status.
Possible values:
Absent, Failure, Good, Simulated (e.g.
when in test mode)
6.2.3 Setup Menu
6.2.3.4.5.1.1 Adjust Conversion
Setup..
In-/Outputs..
Analog Inputs..
Analog Input1 ... 2
Adjust Conversion..