8Troubleshooting
8.2.1 Analyzer Related Messages
8.2.1 Analyzer Related NAMUR Messages
Message
Status level Explanation Recommended Actions
Calculator program error
Maintenance
While running the calculator,
an inconsistency was detec-
ted
• Check the calculator pro-
gram for syntax errors, im-
possible commands or signal
references
• Check for divisions by 0
Division by 0 detected
Can´t open Data Logger
le
Maintenance
Accessing the data logger le
is not possible • Check if internal disk is pre-
sent / installed
• Call Emerson Service
Can´t write to Data Logger
le
Maintenance
Accessing the data logger le
is not possible
Checksum error
Maintenance
Creating the factory congu-
ration le caused a checksum
error.
Store a new factory congura-
tion le.
Cfg checksum error
Maintenance
Creating the user congura-
tion le caused a checksum
error.
Create a new user congura-
tion le.
Cfg le open error
Maintenance
Opening the user congurati-
on le is not possible.
Call Emerson service
Cfg le read error
Maintenance
Writing to the user congurati-
on le is not possible.
Cfg le write error
Maintenance
Writing to the user congura-
tion le is not possible
Device not in Service
Function check
Operator has set the analyzer
to function check mode Set analyzer into service
DISK Free space warning
Maintenance
Internal disk usage exceeded
specied limit (default: 80 %)
Free space by deleting les
DISK full
Maintenance
Internal disk usage exceeded
95 % limit