Emerson Process Management GmbH & Co. OHG 7-23
X-STREAM XE
Instruction Manual
HASXEE-IM-HS
10/2010
7Maintenance & Procedures
7.4.3 Advanced Calibration
start to
zero all channels
with the current
zero gas valve
assigned, conside-
ring the individual
purge times
select
channel 1
increase
channel no.
by 1
wait until
all selected
channels are
zeroed
select all
channels with
the same
zero gas valve
assigned
is current
channel the
last channel?
is this
channel
already
zeroed?
start
zero all
procedure
END
Yes
Yes
No
No
Fig. 7-4: Zero All Calibration Procedure Flow Diagram