Emerson Process Management GmbH & Co. OHG9-36
X-STREAM XE
Instruction Manual
HASXEE-IM-HS
04/2010
9.4 List of Parameters and Registers - Sorted by Daniel Registers
Address Tag Name Data
Type
Client
Access Description
Daniel Modicon
3281 3281 Setup.In/Outputs.AO.SignalAsgn3 Word R/W assigned signal output3
3282 3282 Setup.In/Outputs.AO.OutRange3 Word R/W range of output3 (0=0..20mA, 1=4-20mA)
3283 3283 Setup.In/Outputs.AO.AutoScale3 Word R/W Auto scale for ranged signals on output3 (0=NO,
1=Yes)
3284 3284 Setup.In/Outputs.AO.FailMode3 Word R/W Behavior on errors output3 (0=Track,
1=-10%Start, 2=+10%End)
3285 3285 Setup.In/Outputs.AO.Hold3 Word R/W Hold output3 on calibrations (0=Track, 1=Hold)
3286 3286 Setup.In/Outputs.AO.SignalAsgn4 Word R/W assigned signal output4
3287 3287 Setup.In/Outputs.AO.OutRange4 Word R/W range of output4 (0=0..20mA, 1=4-20mA)
3288 3288 Setup.In/Outputs.AO.AutoScale4 Word R/W Auto scale for ranged signals on output4 (0=NO,
1=Yes)
3289 3289 Setup.In/Outputs.AO.FailMode4 Word R/W Behavior on errors output4 (0=Track,
1=-10%Start, 2=+10%End)
3290 3290 Setup.In/Outputs.AO.Hold4 Word R/W Hold output4 on calibrations (0=Track, 1=Hold)
3291 3291 Setup.In/Outputs.AO.SignalAsgn5 Word R/W assigned signal output5
3292 3292 Setup.In/Outputs.AO.OutRange5 Word R/W range of output5 (0=0..20mA, 1=4-20mA)
3293 3293 Setup.In/Outputs.AO.AutoScale5 Word R/W Auto scale for ranged signals on output5 (0=NO,
1=Yes)
3294 3294 Setup.In/Outputs.AO.FailMode5 Word R/W Behavior on errors output5 (0=Track,
1=-10%Start, 2=+10%End)
3295 3295 Setup.In/Outputs.AO.Hold5 Word R/W Hold output5 on calibrations (0=Track, 1=Hold)
3301 3301 Setup.Calibration.ZeroMethod1 Word R/W zero cal method comp1 (0=Instant, 1=Stability)
3302 3302 Setup.Calibration.ZeroMethod2 Word R/W zero cal method comp2 (0=Instant, 1=Stability)
3303 3303 Setup.Calibration.ZeroMethod3 Word R/W zero cal method comp3 (0=Instant, 1=Stability)
3304 3304 Setup.Calibration.ZeroMethod4 Word R/W zero cal method comp4 (0=Instant, 1=Stability)
3305 3305 Setup.Calibration.ZeroMethod5 Word R/W zero cal method comp5 (0=Instant, 1=Stability)
3306 3306 Setup.Calibration.SpanMethod1 Word R/W span cal method comp1 (0=Instant, 1=Stability)
3307 3307 Setup.Calibration.SpanMethod2 Word R/W span cal method comp2 (0=Instant, 1=Stability)
3308 3308 Setup.Calibration.SpanMethod3 Word R/W span cal method comp2 (0=Instant, 1=Stability)
3309 3309 Setup.Calibration.SpanMethod4 Word R/W span cal method comp4 (0=Instant, 1=Stability)
3310 3310 Setup.Calibration.SpanMethod5 Word R/W span cal method comp5 (0=Instant, 1=Stability)
3397 3397 Setup.Communication.SIntMod-
busFt32 Word R/W Modbus 32Bit mode of serial COM
3398 3398 Setup.Communication.SSvcMod-
busFt32 Word R/W Modbus 32Bit mode of serial service COM
3399 3399 Setup.Communication.Eth2Mod-
busFt32 Word R/W Modbus 32Bit mode of Ethernet2
3400 3400 Setup.Communication.Eth1Mod-
busFt32 Word R/W Modbus 32Bit mode of Ethernet1
3401 3401 Service.Measurement.Compensati-
on.ToffEnable1 Word R/W enable temperature zero compensation of comp1
3402 3402 Service.Measurement.Compensati-
on.ToffEnable2 Word R/W enable temperature zero compensation of comp2
3403 3403 Service.Measurement.Compensati-
on.ToffEnable3 Word R/W enable temperature zero compensation of comp3