Instruction Manual
Appendix D Rev. 2.4
July 1998
D-2 Appendices Rosemount Analytical Inc. A Division of Emerson Process Management
World Class 3000
POWER
SUPPLY
REGULATOR
TUBE
SOLENOID MANIFOLD HOSE
ADAPTER
CABLE
GRIP
TERMINATION
BOARD
FLOWMETER
27270024
Figure D-2. Multiprobe Calibration Gas Sequencer, Interior
Table D-1. Specifications for Multiprobe Calibration Gas Sequencer.
Environmental Classification ................................................................ NEMA 4X (IP56)
Humidity Range ....................................................................................95% Relative Hum idity
Humidity Range ....................................................................................95% Relative Hum idity
Ambient Temperature Range ............................................................... -20° to 160°F (-30° to 71°C)
Vibration................................................................................................ 5 m/sec2, 10 to 500 xyz plane
External Electrical Noise....................................................................... Minimum Interference
Piping Distance Between MPS 3000 and Probe .................................. Maximum 300 feet (91 m)
Cabling Distance Between MPS 3000 and Electronic s Package.........Max imum 1000 feet (303 m)
In Calibration Status Relay ...................................................................48V max , 100 mA max
Cabling Distance Between MPS 3000 and Status Re lay Indicator...... Maximum 1000 feet (303 m)
Approximate Shipping Weight .............................................................. 35 pounds (16 kg)