6-4 HEATER PROBLEM
For all heater troubleshooting, allow at least 30
minutes for the operating temperature to stabi-
lize. After the warmup period, observe the s ys-
tem status and the voltages of the cell TC and
the cold junction AD590. For heater related
problems:
a. The status line may read: HtrEr or OFF.
b. The displayed O2 value will read 0%.
c. Cell TC voltages will vary from normal.
These voltages are found by accessing the
proper menu. In the IFT, use the DIAG-
NOSTIC DATA sub-menu of the PROBE
DATA menu.
Refer to Table 6-2 to troubleshoot heater related
problems.
Table 6-2. Heater Troubleshooting
Problem
Cause
Corrective Action
Status is HtrEr or OFF.
Cell TC < 28.4 mV.
Cold Junction 273 to 330 mV (normal).
O2 Display = 0%
1. Blown fuse or faulty wiring.
Check all fuses and wiring for continuity and re pair as needed. Verify that input power
jumpers are installed correctly. Check jumper s for proper configuration in IFT and HPS if used.
2. Heater failure.
In HPS with power OFF, check heater resistance at J2, term inals R/H. For 44 V heater, resis-
tance should be 11 to 14 ohms. For 115 V heater, resis tance should be 67 to 77 ohms. Check
wiring, and replace heater if needed. Heater resistanc e can also be checked at the probe junc-
tion box:
44 V heater: terminals 7 and 8 should measure 11 to 14 ohms.
115 V heater: terminals 5 and 6 should measure 67 to 77 ohms. (Terminals 6 to 7 and 6 to
8 should be open circuits.)
3. Triac open.
Check the triac. Repair as needed.
4. Electronics failure.
First check and repair all related wiring. Check and repair electronics as needed.
5. Missing insulation around heater.
Check that insulation is in place and undamaged. R epair or replace insulation as needed.