World Class 3000

Instruction Manual

IB-106-300NH Rev. 4.2 July 2002

Table 5-5. SETUP Sub-Menu (continued)

SUB-MENU

 

 

 

 

 

 

 

SELECTION

 

PARAMETERS

 

DESCRIPTION

 

 

Relay

 

 

 

 

 

 

 

 

Outputs

 

 

 

 

 

 

 

 

 

 

 

 

 

NOTE

 

 

 

K1 and K2 relay outputs can be configured for "OFF" or any one of the eight

 

 

 

events listed below. Up to three events can control each relay output. Events are

 

 

 

selected in the SETUP sub-menu.

 

 

 

 

K1 Setup

 

 

-

Off

 

No effect.

 

 

 

 

Event 1

1.

In Cal

 

Probe goes into calibration status.

 

 

 

 

Event 2

2.

Hi O2

 

Output exceeds high end alarm limit.

 

 

 

 

Event 3

3.

Lo O2

 

Output goes below low alarm limit.

 

 

 

 

Event 1

4.

Htr Fail

 

Probe heater fault occurs.

 

 

K2 Setup

 

Event 2

5.

Cal Fail

 

Probe failed last calibration.

 

 

 

 

Event 3

6.

TG Low

 

Calibration gas pressure gets too low.

 

 

 

 

 

7.

Cell Res

 

Probe resistance exceeds high limit.

 

 

 

 

 

8.

High Range

 

High analog output range is selected.

 

 

 

 

 

 

 

 

Analog

 

SOURCE

O2

 

 

Select the measurement value to be represented on

 

 

Output

 

 

Efficiency

 

the analog output.

 

5

 

 

 

Dual Rng O2

 

 

 

 

 

 

 

 

 

 

AOUT TYPE

 

 

 

 

 

 

 

 

 

Select one of the listed options to define upper and

 

 

 

 

 

HART 4-20mA

 

lower limits of probe analog output. Only a selection

 

 

 

 

 

0-20mA

 

that matches the position of the analog output selector

 

 

 

 

 

0-10V

 

switch on the microprocessor board (Figure 2-9)will

 

 

 

 

 

 

 

 

be accepted. The defined limits correspond to the

 

 

 

 

 

 

 

 

upper-lower %O2 values defined in the Range Setup

 

 

 

 

 

 

 

 

menu.

 

 

 

 

RANGE SETUP

 

 

 

 

 

 

 

(Source not set to Dual Rng O2)

 

 

 

 

 

 

Xfer Fnct

Log

 

 

Select the transfer function used on the analog output.

 

 

 

 

 

Lin

 

 

Selecting Log will not effect the output when Efficiency

 

 

 

 

 

 

 

 

is selected as the Source.

 

 

 

 

Range Values

 

 

 

 

 

 

 

High End

 

 

Enter the upper and lower analog output range values.

 

 

 

 

0.000% O2 - 25.00% O2

 

The High End value defines the measured O2 value

 

 

 

 

Low End

 

 

corresponding to the high analog output value, i.e.,

 

 

 

 

 

 

20mA or 10V, and the Low End value corresponds to

 

 

 

 

0.000% O2 - 25.00% O2

 

 

 

 

 

 

the low analog output value, i.e., 0mA, 4mA, or 0V.

 

 

 

 

 

 

 

 

 

 

Rosemount Analytical Inc. A Division of Emerson Process Management

General User Interface (GUI) Operation 5-13

Page 83
Image 83
Emerson Process Management 3000 instruction manual Xfer Fnct, Range Values

3000 specifications

Emerson Process Management is a key player in the field of industrial automation, and one of its flagship offerings is the Emerson Process Management 3000. This system is designed to optimize production processes across a variety of industries, enhancing efficiency and ensuring robust operational performance.

One of the main features of the Emerson Process Management 3000 is its advanced control capabilities. The system employs model predictive control (MPC) technology, which allows for real-time data processing and decision-making. By predicting future trends in process variables, the MPC enables operators to adjust control parameters proactively, minimizing fluctuations and improving overall productivity.

Additionally, the Emerson 3000 integrates seamlessly with existing industrial infrastructure. This system supports a wide range of communication protocols, ensuring compatibility across various devices and sensors. Its flexibility allows companies to leverage their investment in existing technologies while upgrading to more modern control strategies.

The user interface of the Emerson Process Management 3000 is highly intuitive, featuring graphical displays that provide operators with clear visibility into process parameters. This interface is designed for ease of use, with customizable dashboards that can be tailored to meet the specific needs of the users. Training time is significantly reduced due to the straightforward design, leading to quicker onboarding for new operators.

In terms of data analytics, the Emerson 3000 also stands out. It features built-in data logging and trend analysis capabilities, empowering operators to make informed decisions based on historical and real-time data. This analytical strength helps in identifying inefficiencies and potential areas for improvement, contributing to reduced downtime and lower operational costs.

Safety is another critical characteristic of the Emerson Process Management 3000. The system includes robust safety features and compliance with various industry standards. By integrating safety into the control process, it ensures that product quality and equipment integrity are maintained, thus safeguarding both personnel and assets.

Overall, Emerson Process Management 3000 is a sophisticated control system that combines advanced technologies, user-friendly design, and robust safety features. Ideal for industries such as oil and gas, chemicals, and water treatment, it is designed to enhance operational efficiency and drive productivity in modern industrial environments.