SECTION 4 CALIBRATION

4-1 ANALOG OUTPUT CALIBRATION

For the analog output to perform within the pub- lished specifications, it must be manually cali- brated. The only equipment needed to perform the calibration is a voltage or current meter, de- pending on which mode of operation is to be calibrated. Prior to manual calibration, remove the IFT from any control loops it may be in.

Prior to manual calibration, the IFT should be removed from any auto- matic control loops. Failure to remove the IFT from control loops prior to calibration may result in faulty equip- ment performance.

Once initiated from the Setup - Analog Outputs menu, the calibration procedure is self guiding.

4-2 SYSTEM CALIBRATION

a.Overview

The primary purpose of an oxygen analyzer is to give an accurate representation of the percentage of O2 in the gas stream. The system should be calibrated periodically to maintain an accuracy which may otherwise be reduced over time due to cell aging. A calibration record sheet is provided at the end of this section to track cell performance.

A requirement for calibration is a set of two accurate calibration gases spanning the oxygen range of most interest. For example, 0.4% and 8% for a 0-10% oxygen range.

Under normal conditions the probe should not need frequent calibration. Because cali- bration is necessary, the system can be equipped with the optional MPS 3000 Multi- probe Calibration Gas Sequencer for fully automatic calibration at regular intervals. Without an MPS, the probes must be cali- brated manually (semiautomatically).

b.Probe Calibration

1.Previous Calibration Constants Functionality

There are three sets of registers used to store calibration constants. These are: Latest Calibration, Previous Cali- bration, and Calculation. Only the val- ues in the Calculation register are used

to calculate the oxygen value for dis-

play and representation on the analog 4 output signal. These values may be

changed in two ways.

(a)The operator may change the val- ues through the SETUP menu. The operator may adjust the slope and constant individually, or reset both to the values calculated during the last good calibration. To reset the values, move the cursor to RESET SLOPE & CONST and push

ENTER.

(b)The IFT will automatically change the values after each calibration as follows:

The values in the Latest Calibra- tion registers are updated after every complete calibration, even if the calibration is not successful. If the calibration is successful, the values in the Latest Calibration registers are copied into the Previ- ous Calibration registers. This is accomplished prior to the update of the Latest Calibration registers. The new slope and constant are copied into the Calculation register.

If the calibration fails, the Previous Calibration registers retain their existing values, while the Latest Calibration registers record the values of the failed calibration. The Calculation register is not updated when the calibration fails.

Rosemount Analytical Inc. A Division of Emerson Process Management

Calibration 4-1

Page 61
Image 61
Emerson Process Management 3000 instruction manual Analog Output Calibration, System Calibration, Overview

3000 specifications

Emerson Process Management is a key player in the field of industrial automation, and one of its flagship offerings is the Emerson Process Management 3000. This system is designed to optimize production processes across a variety of industries, enhancing efficiency and ensuring robust operational performance.

One of the main features of the Emerson Process Management 3000 is its advanced control capabilities. The system employs model predictive control (MPC) technology, which allows for real-time data processing and decision-making. By predicting future trends in process variables, the MPC enables operators to adjust control parameters proactively, minimizing fluctuations and improving overall productivity.

Additionally, the Emerson 3000 integrates seamlessly with existing industrial infrastructure. This system supports a wide range of communication protocols, ensuring compatibility across various devices and sensors. Its flexibility allows companies to leverage their investment in existing technologies while upgrading to more modern control strategies.

The user interface of the Emerson Process Management 3000 is highly intuitive, featuring graphical displays that provide operators with clear visibility into process parameters. This interface is designed for ease of use, with customizable dashboards that can be tailored to meet the specific needs of the users. Training time is significantly reduced due to the straightforward design, leading to quicker onboarding for new operators.

In terms of data analytics, the Emerson 3000 also stands out. It features built-in data logging and trend analysis capabilities, empowering operators to make informed decisions based on historical and real-time data. This analytical strength helps in identifying inefficiencies and potential areas for improvement, contributing to reduced downtime and lower operational costs.

Safety is another critical characteristic of the Emerson Process Management 3000. The system includes robust safety features and compliance with various industry standards. By integrating safety into the control process, it ensures that product quality and equipment integrity are maintained, thus safeguarding both personnel and assets.

Overall, Emerson Process Management 3000 is a sophisticated control system that combines advanced technologies, user-friendly design, and robust safety features. Ideal for industries such as oil and gas, chemicals, and water treatment, it is designed to enhance operational efficiency and drive productivity in modern industrial environments.