Instruction Manual
IB-106-300NH Rev. 4.2
July 2002
5-10 General User Interface (GUI) Operation Rosemount Analytical Inc. A Division of Emerson Process Management
World Class 3000
Table 5-4. CALIBRATE O2 Sub-Menu
SUB-MENU
SELECTION
SETUP SETTING
(SEE TABLE 3-5) DISPLAY DESCRIPTION
Perform
Calibration
Auto Cal in Probe
Setup is YES
Press ENTER to start Auto
Calibration.
MPS will start calibrating probe.
Starting Automatic Calibration
High Gas _____% O2
Time Left 0:00
Value for high O2 calibration gas.
Amount of time necessary to com-
plete the current testing phase in
min:sec.
Cell mV ______mV Cell voltage of the probe.
Low Gas _____% O2
Time Left 0:00
Value for low O2 calibration gas.
Amount of time necessary to com-
plete the current testing phase in
min:sec.
Cell mV ______mV Cell voltage of the probe.
Resistance Check
Time Left 0:00
Resistance check in progress.
Cell _____mV _____C
Calibration Complete
Purging 0:00
Cell voltage and probe temperature.
Gas lines are being purged of cali-
bration gas.
Cell _____mV _____C
Calibration Complete
Cell voltage and probe temperature.
Auto Cal in Probe
Setup is NO.
Press ENTER to start Manual
Calibration.
Manual calibration sequence will
begin when ENTER is pressed.
Switch ON high calibration gas.
Press ENTER when ready.
High gas ______% O2High O2 calibration gas value .
Press ENTER when O2 reading
is stable.
Turn OFF high calibration gas
and ON low calibration gas.
Press ENTER when ready.
Low gas ______% O2Low O2 calibration gas value.
Press ENTER when O2 reading
is stable.
Resistance Check. Resistance check in progress.
Turn off low calibration gas.
Press ENTER when ready.
Press ENTER when probe has
returned to process.