Installing the Dampening Material, Transducers and Fixtures 3-1
April 2004

Chapter 3

Installing the Dampening

Material, Transducers and

Fixtures

Since the PT878GC is specifically de signed for gas m easur ement
with clamp-on transducers, it requires the use of specially
designed fixtures and dampening material to maintain the highest
possible measurement accuracy. GE Panametrics supplies the
CFG series of fixtures:
The V1 clamping fixture for pipes with diameters between
0.75 and 1.25 in. (20 to 30 mm).
The V4 clamping fixture for pipes with diameters between
1.25 and 4 in. (30 to 100 m m).
The V8 clamping fixture for pipes from 4 to 8 in. (100 to
200 mm).
The V12 clamping fixture for pipes from 8 to 12 in. (200 to
300 mm).
The PI clamping fixture for pipes from 12 to 24 in. (300 to
600 mm).
Figure 3-1 on the next page illu strates the V series and PI fixture
assemblies. Complete the steps in the following sections to
position and install the transducers, fixtures and dampening
material.