Installing the PI Fixture and Transducers

The PI clamping fixture holds transducers on pipes from 8 to 24
in. in diameter. It comes with either a chain or strap, depending
on the selection made with the initial order from GE Panametrics.
To install the fixture and transducers, complete the following
steps:

Surveying the Pipe

1. Measure the pipe circumference to an accuracy of ±2 mm
(±1/16 in.)
IMPORTANT: Do not use a calculated value or a nominal value
for the circumference.
2. Tightly wrap the layout tape once around the entire pipe and
line up the edges. Using the tape as a template guide for
marks, mark scribe lines around the entire diamet er of the
pipe, as shown in Figure 3-13 below.
Figure 3-13: Marking Circumferential Lines on the Pipe
3. Line up the zero scale of the layout tape at the desired location
of the first transducer. (For a typical installation, this point
will be the 3 o’clock position on a horizontal pipe.) Mark each
of the two circumferential lines at the zero point. Connect
each of these marks using a straight edge (for example, the
edge of the layout tape) as sh own in Figure 3-14 on the next
page.