Gilson 215 manual Exyz, Jccccp, Jcttt

Models: 215

1 112
Download 112 pages 10.16 Kb
Page 103
Image 103

Appendix

 

C

Exyz

B

Sets X, Y, and Z motor status:

x

-

0 for disable X motor.

y

-

0 for disable Y motor.

z

-

0 for disable or 1 for enable Z motor.

 

 

For example, the following command disables the motors:

 

 

E000.

 

 

Fn

B

Relaxes the probes where “n” is the probe selection (a, b, c,

 

 

or d).

 

 

H

B

Moves XYZ to home position, homes all four Z-drives, and

 

 

sets the X-pitch to 18 mm.

I

I

Reads status of input contacts and front panel Start and

 

 

Stop buttons. Returns “cccckk” where:

 

 

cccc

-

Status of input contacts A, B, C, and D: 1 if

 

 

 

 

closed (shorted), 0 if open.

 

 

kk

-

Status of front panel Start and Stop buttons: 1 if

 

 

 

 

button has been pressed, 0 if it hasn’t been

 

 

 

 

pressed. Value of kk is cleared after reading.

J

I

Reads status of output contacts. Returns “ccccp” where:

cccc- Status of output contacts 1, 2, 3, and 4: 1 if connected, 0 if disconnected.

p- Status of keypad and auxiliary power: 1 if connected, 0 if disconnected.

Jcccc[p]

B

Sets output contacts and +24V external power.

cccc- Output contacts 1, 2, 3, and 4: 1 to connect, 0 to disconnect, X for no change.

p- (Optional) Keypad and auxiliary power: 1 to connect, 0 to disconnect, X for no change.

jcttt

B

Pulses an output contact:

c - Number of the output contact, 1–4.

t- Duration of the pulse in tenths of seconds; default is 1 which is 100 ms.

GSIOC Commands

C-5

Page 103
Image 103
Gilson 215 manual Exyz, Jccccp, Jcttt