Maintenance

Replacing Parts

4

Installing different type of probe

To install a replacement probe of a different type than is currently installed, you may want to obtain a new probe guide insert for precise XY probe accuracy.

1Remove the transfer tubing’s 1/4"-28 fitting connected to the top of the isolation probe holder.

2Grasp the current probe and push it up through the top of the isolation probe holder.

3Remove the current probe guide insert by turning it counterclockwise. Then place the new probe guide insert into probe guide and secure it by turning it clockwise.

4Insert the new probe into the top of the isolation probe holder and pull it through the holder and the new probe guide insert until the tip of the probe is in the probe guide.

5Replace and tighten the 1/4"-28 fitting.

4-10

Page 66
Image 66
Gilson 215 manual Installing different type of probe