Maintenance

Replacing a Level Sensing Cable

Removing the level sensing cable

1Detach the transfer tubing from the top of the isolation probe holder by using the supplied headless nut extender (part number 49041032) to loosen and remove the 1/4"-28 nut and ferrule.

2Unplug the right-angle connector for the liquid level sensing cable (LLD cable) from the socket on the back of the Z-arm.

3Detach the LLD cable from the LLD cable guide.

4Loosen the small hexagonal nut on the probe holder for probe one. Detach the split- tongue connector from the nut.

5Remove the transfer tubing from the LLD cable casing.