Appendix

GSIOC Commands

Ksxxx

B

K

I

 

MI

mI

MB

mB

C

Sets liquid level sensing sensitivity where:

s - selected probe (a, b, c, or d)

xxx- Value from 0 to 255; 0 is the most sensitive, 255 is the least sensitive.

If s is not specified, the default is abcd.

Reads liquid level sensing sensitivity. Returns “aaa,bbb,ccc,ddd” where:

aaa- sensitivity of probe A

bbb- sensitivity of probe B

ccc- sensitivity of probe C

ddd- sensitivity of probe D

Possible sensitivity range is from 0 to 255 for each probe; 0 is the most sensitive, 255 is the least sensitive.

Reads X, Y, Z motor status and dilutor status. Returns “xyzp”. For each status, you’ll see U for unpowered, P for powered, R for running, or E for error.

Note: Because the Quad-Z uses an external dilutor, the dilutor status is always unpowered (U).

Read X, Y, Z motor status and width of the probes. Returns “xyabcdw” where:

x = X motor y = Y motor

a = Z motor on probe A b = Z motor on probe B c = Z motor on probe C d = Z motor on probe D w = pitch motor

For each status, you’ll see U for unpowered, P for powered, R for running, or E for error.

Start to move probes to the Z-height positions (set using the buffered T command, see page C-8).

Start to move probes to the Z-height positions (set using the buffered T command, see page C-8)with liquid level sensing on.

C-6

Page 104
Image 104
Gilson 215 manual Ksxxx