Appendix
GSIOC Commands | Ksxxx | B | |
K | I | ||
|
MI
mI
MB
mB
C
Sets liquid level sensing sensitivity where:
s - selected probe (a, b, c, or d)
xxx- Value from 0 to 255; 0 is the most sensitive, 255 is the least sensitive.
If s is not specified, the default is abcd.
Reads liquid level sensing sensitivity. Returns “aaa,bbb,ccc,ddd” where:
aaa- sensitivity of probe A
bbb- sensitivity of probe B
ccc- sensitivity of probe C
ddd- sensitivity of probe D
Possible sensitivity range is from 0 to 255 for each probe; 0 is the most sensitive, 255 is the least sensitive.
Reads X, Y, Z motor status and dilutor status. Returns “xyzp”. For each status, you’ll see U for unpowered, P for powered, R for running, or E for error.
Note: Because the
Read X, Y, Z motor status and width of the probes. Returns “xyabcdw” where:
x = X motor y = Y motor
a = Z motor on probe A b = Z motor on probe B c = Z motor on probe C d = Z motor on probe D w = pitch motor
For each status, you’ll see U for unpowered, P for powered, R for running, or E for error.
Start to move probes to the