Operation

Configuring the Liquid Handler

3

Adjust XY

The Adjust XY tab allows you to test whether the instrument is properly adjusted and to make minor adjustments to the X- and Y-axis offsets if needed. You may need to use the options under this tab if the probe is not accessing the injection port of the injection module installed or the vessels in the installed racks.

The X offset and Y offset text boxes display the current offsets stored in the instrument’s memory.

To determine if the probe(s) need(s) to be adjusted in the X or Y direction select the model of the injection module that is on your Quad-Z 215 or select other for a user-defined test point. The default XY coordinates are shown next to the model number.

Select the model of the injection module that is on your 215 or select other for a user-defined test point. The default XY coordinates are shown next to the model number.

3-10