Reads the liquid detector output. Returns “a,b,c,d” where:

 

 

a - leftmost probe (probe A)

 

 

b - probe B

 

 

c - probe C

 

 

d - probe D

 

 

For each status, you’ll see A for air or L for liquid.

Oabcd

B

Sets each probe speed value (in µm/s) where:

 

 

a = speed of probe A

 

 

b = speed of probe B

 

 

c = speed of probe C

 

 

d = speed of probe D

 

 

If the value is not specified, the probe speed value will

 

 

remain the same.

 

 

The default speed is 125000 µm/s.

O

I

Reads each probe speed value (in µm/s). Returns “a,b,c,d”

 

 

where:

 

 

a = speed of probe A

 

 

b = speed of probe B

 

 

c = speed of probe C

 

 

d = speed of probe D

P

I

Reads X and Y encoder positions in integral units of 0.1 mm.

q

I

Reads probe range values in the X direction for each probe

 

 

(a–d) in sequence where:

 

 

a = minimum X value/maximum X value

 

 

b = minimum X value/maximum X value

 

 

c = minimum X value/maximum X value

 

 

d = minimum X value/maximum X value

 

 

The range is displayed in integral units of 0.1 mm. The

 

 

values are based on the current probe pitch. Changing the

 

 

pitch will alter the ranges.