Appendix |
| C |
N | I | Reads the liquid detector output. Returns “a,b,c,d” where: |
|
| a - leftmost probe (probe A) |
|
| b - probe B |
|
| c - probe C |
|
| d - probe D |
|
| For each status, you’ll see A for air or L for liquid. |
Oabcd | B | Sets each probe speed value (in µm/s) where: |
|
| a = speed of probe A |
|
| b = speed of probe B |
|
| c = speed of probe C |
|
| d = speed of probe D |
|
| If the value is not specified, the probe speed value will |
|
| remain the same. |
|
| The default speed is 125000 µm/s. |
O | I | Reads each probe speed value (in µm/s). Returns “a,b,c,d” |
|
| where: |
|
| a = speed of probe A |
|
| b = speed of probe B |
|
| c = speed of probe C |
|
| d = speed of probe D |
P | I | Reads X and Y encoder positions in integral units of 0.1 mm. |
q | I | Reads probe range values in the X direction for each probe |
|
| |
|
| a = minimum X value/maximum X value |
|
| b = minimum X value/maximum X value |
|
| c = minimum X value/maximum X value |
|
| d = minimum X value/maximum X value |
|
| The range is displayed in integral units of 0.1 mm. The |
|
| values are based on the current probe pitch. Changing the |
|
| pitch will alter the ranges. |
GSIOC Commands