Installation

Probe Installation

2

Probe Installation

There are different probes available for use on the Quad-Z 215 Liquid Handler. Depending upon your application, you have purchased the appropriate probes and probe guide inserts. When installing the probes or custom sized probe guide inserts (1.3 or

1.5mm) refer to the following procedures and diagrams.

Installing the Probe Guide Inserts

Your Quad-Z is delivered with 1.5 mm probe guide inserts installed. If you ordered custom sized probe guide inserts for more precise XY accuracy, refer to the installation instruction below.

To install the probe guide inserts, turn the insert clockwise into the probe guide.

Installing the Probes

Insert the probes into the top of the isolation probe holders and pull them through the holders and the probe guide inserts until the tip of the probe is in the probe guides.