Installation

Electrical Connections

2

Baud rate/mode

As a default, the baud rate/mode is set to 6, indicating that the liquid handler is set for a baud rate of 19200 and is a master device.

You’ll need to change the selection to 0 (zero) if the liquid handler is connected via the GSIOC to a Gilson system and is being controlled by the 506C System Interface from Gilson control software. A setting of 0 indicates the liquid handler is a slave device and the baud rate is being clocked externally.

Note that other selections are available if the liquid handler is being controlled by non-Gilson applications. Refer to the table shown on the liquid handler’s rear panel; see page 2-16.

To change the baud rate/mode:

1Gently insert a small flat blade screwdriver into the SW2 selector on the rear panel and turn it.

2Align the white dot with one of the indicated numbers.

2-22

Page 38
Image 38
Gilson 215 manual Baud rate/mode