Troubleshooting

Liquid Level Detector

Liquid level detector not detecting liquid level

Ensure that the level sensing cables are plugged in.

Check sensitivity setting in the 215 Setup Utility (see page 3-9) and lower the percentage.

Check if liquid is detectable. Liquid level detection works only if there is electrical conductivity in your liquid. Liquid level detecting will not work with most non-polar liquids. For intermediate polarity liquids and polar liquids, check the sensitivity setting in the 215 Setup Utility.

Call Gilson if this is caused by faulty circuitry.

Liquid level detector falsely detecting liquid level

Ensure that probes are installed correctly.

Check sensitivity setting in the 215 Setup Utility (see page 3-9)and raise the percentage.

Call Gilson if this is caused by faulty circuitry.