Installation

4Loosen the mounting screw on the Z-arm mounting bracket and slightly raise the Z-arm.

5Manually move the liquid handler’s arm so the Z-arm is centered over the Z-height adjustment tool.

6While holding Z-height adjustment tool flat against the locator plate, use the other hand to lower the Z-arm until it lightly rests on the adjustment tool.

7Tighten the mounting screw on the Z-arm mounting bracket so the Z-arm is secure.

8While holding the adjustment tool in place, slide the Z-arm off the tool. Ensure that the bottom of the Z-arm lightly rubs against the adjustment tool as it moves. Repeat steps 4 through 7 until this is true.

9Store the Z-height adjustment tool.

2

Final Z-ArmAdjustment