Maintenance
Acidic and basic compounds - Follow this procedure if the fluid path is in contact with acidic and basic compounds:
1Prime the fluid path using a 0.1N NaOH solution.
2Pause the priming sequence.
3After 10 minutes, resume priming the fluid path using distilled or deionized water. Prime until the fluid path has been flushed with water.
4Pause the priming sequence.
5Prime the fluid path using a 0.1N NaOH solution. Continue to prime until the fluid path has been flushed with 0.1N NaOH.
6Pause the priming sequence.
7After 10 minutes, resume priming the fluid path using distilled or deionized water. Prime until the fluid path has been flushed with water.
8When you’re satisfied that the entire fluid path has been flushed with water, end the priming sequence.