Troubleshooting

5

 

 

 

Error Messages

Error Messages

Error

Description

Solution

15

NV-RAM checksum is invalid

• Send the buffered ~9 GSIOC command to

 

 

reset the NV-RAM and initialize to the

 

 

default value. Run 215SETUP.EXE or the

 

 

215 Setup Utility program

 

 

• Replace the main board

16

X scale factor is invalid

Contact the Gilson Customer Service Department.

17

Y scale factor is invalid

Contact the Gilson Customer Service Department.

20

X motor position error

Turn power off then on to the liquid handler.

21

Y motor position error

Turn power off then on to the liquid handler.

22

Z motor position error

Turn power off then on to the liquid handler.

24X target less than minimum X Send the immediate Q command using the GSIOC Utility Program to read the travel range. Correct the error in the program controlling the liquid handler.

25X target more than maximum X Send the immediate Q command using the GSIOC Utility Program to read the travel range. Correct the error in the program controlling the liquid handler.

26Y target less than minimum Y Send the immediate Q command using the GSIOC Utility Program to read the travel range. Correct the error in the program controlling the liquid handler.

27Y target more than maximum Y Send the immediate Q command using the GSIOC Utility Program to read the travel range. Correct the error in the program controlling the liquid handler.

28Z target less than minimum Z Send the immediate Q command using the GSIOC Utility Program to read the travel range. Correct the error in the program controlling the liquid handler.