7.6 Re-installing the Bottom Panel

Section 7

To open the controller:

1.Remove the 17 screws from around the perimeter of the controller back panel.

2.Carefully pull open the bottom panel and let the attached connectors swing out of the way.

3.If necessary for the repair being performed, unplug the appropriate connectors. Always note each connector location before unplugging anything. Re-installing a plug into the wrong connector can cause extensive damage.

Note: The front panel gasket has a light coating of grease to help ensure a watertight

seal. Do not contaminate the grease or gasket area with dirt or debris during servicing. Always replace the gasket if it is damaged or missing. Never reassemble the case without the gasket properly installed.

7.6 Re-installing the Bottom Panel

Note: Always follow the procedure below when re-installing the sampler bottom panel. Improper panel installation may result in damage to

the instrument.

1.Hand tighten each nut until it makes contact with the panel.

2.Tighten the nuts in sequence shown in Figure 30 to 5 in.-lb (0.56 N-m).

3.Repeat the tightening procedure in the same sequence to 10 in.-lb (1.13 N-m).

Figure 30 Tightening the Nuts on the Bottom Panel

1 10

9Figure 30 Tightening the Nuts on the Bottom Panel

178

8 2 11

16 Manual backgroundManual background

7 15

2 11

Manual background 3

Manual background 12

4

Manual background 13 Manual backgroundManual background 5

6 14

Page 100

 

Re-installing the Bottom Panel

8990mnt.fm

Page 102
Image 102
Hach 900 MAX manual 8 16 7, 2 3 12 4 13 6, Re-installingthe Bottom Panel, Section, Tightening the Nuts on the Bottom Panel