Trouble Alarm Conditions, Causes, and Solutions

Appendix C

Trouble Alarm Conditions, Causes, and Solutions

Trouble Condition

Cause

Solution

 

 

 

Low Main Battery

Power supply voltage is less than 11 V dc.

Change batteries.

 

 

 

Memory Battery

Internal memory battery voltage is too low.

Change memory batteries.

 

 

 

 

Free slate memory is less than 20%. RAM

Download data from unit, halt and restart the

Low Slate Memory

memory is almost full and will stop

program or download data, halt, and change

 

recording soon.

data to wrap mode.

 

 

 

 

No more slate memory. Unit is in slate memory

Download data and restart the program or

Slate Memory Full

download data then change the

mode and cannot log any more data.

 

memory mode.

 

 

 

 

 

 

 

If the modem is enabled, the problem may be

Modem Failure

Unable to initialize modem PCB.

found on the modem board and the board

 

 

may need to be replaced.

 

 

 

Missed Sample

Sample cycle did not result in getting a sample.

 

 

 

Purge Failure

Sensor(s) still detecting liquid after a line purge.

 

 

 

Jammed Distributor

Distributor arm has jammed on an obstruction.

Remove obstruction

 

 

 

 

Refrigeration system is unable to reduce

 

Unable to Cool

temperature in refrigeration compartment to

 

user defined setpoint.

 

 

 

 

 

Heater is unable to increase temperature in

 

Unable to Heat

refrigeration compartment to user defined

 

setpoint.

 

 

 

 

 

The echo has been temporarily deflected by a

Check for excessive foam. Remove any debris

U-Sonic Echo Loss

change in site conditions such as floating debris

or build-up.

 

or foam in the channel, wind, etc.

 

 

 

 

 

Xducer Ringing

Transducer is operating within the deadband

 

 

 

U-Sonic Failure

Transducer not plugged in. Cable damaged.

Check that the transducer is plugged in and

Transducer thermal sensor damaged.

that there is no damage to the sensor.

 

 

 

 

 

Communication error between CPU and remote

 

ultrasonic module.

 

 

 

 

 

 

CPU is not communicating with the

Wait a few minutes to see if it goes away, if it

 

continues it is an indication that there is a

 

Velocity board.

 

problem with the CPU board.

RS485 Timed Out

 

 

 

 

If the logging interval is 1 or 2 minutes and

 

 

 

 

conditions are poor with the problem

 

Not receiving a good velocity reading.

continuing indefinitely, try increasing the

 

 

logging interval to allow more time to capture

 

 

the signal.

 

 

 

 