Hach 900 MAX manual Specifications

Models: 900 MAX

1 160
Download 160 pages 21.13 Kb
Page 11
Image 11

 

 

Specifications

 

 

Specifications are subject to change without notice.

 

 

 

General

 

 

 

 

 

Dimensions

Width 61 cm (24 in.), Depth 61 cm (24 in.), Height 110 cm (43.5 in.), Weight 63.3 kg (140 lb)

 

 

 

 

Cabinet

Fiberglass reinforced plastic with beige UV inhibited polymer laminate.

 

 

 

 

Temperature Range

-29 to +50 °C (-20 to 122 °F); With optional controller compartment heater, -40 to +50 °C

 

(-40 to 122 °F).

 

 

 

 

 

 

Recovery Time

With door open one minute in 24 °C (75 °F) ambient and 4 °C sample temperature, 5

 

minutes.

 

 

 

 

 

 

Pull-Down Time

From 24 °C (75 °F) to 4 °C (39 °F), 15 minutes.

 

 

 

 

 

Top mounted compressor/condenser with fan forced air cooled condenser; three-sided

 

Thermal System

wrap-around evaporator plate; 2 in. rigid foam insulation; microprocessor controlled

 

thermostat maintains sample liquid at 4 °C (±1 °C); frost free; compression gasket door seal;

 

 

 

 

refrigeration components and plumbing are corrosion protected with conformal coating.

 

 

 

 

Power Requirements

115 V ac, 60 Hz (230 and 100 V ac optional); Compressor Running Amperage 1.5–2.0 A.

 

Locked rotor current 12 amps. Installation Category II.

 

 

 

 

 

 

Optional ac

Pump/Controller Only: Rechargeable 6 amp-hour gel lead acid battery takes over

 

Power Backup

automatically with ac line power failure. Integral trickle charger maintains battery at full

 

(Pump Controller Only)

charge (factory installed option).

 

 

 

 

Internal Battery

Two C cell alkaline batteries; maintains program logic and real time clock for five years.

 

Internal battery current draws less than 40 micro amps.

 

 

 

 

 

 

Control Panel

21 key membrane switch keypad with 4 multiple function soft keys; 8 line x 40 character

 

alphanumeric, back lit liquid crystal graphics display. Self prompting/menu driven program.

 

 

 

 

 

 

 

Records program start time and date, stores up to 400 sample collection times/dates, all

 

 

program entries, operational status including number of minutes or pulses to next sample,

 

 

bottle number, number of samples collected, number remaining, sample identification

 

Data Logging

number, and all logged data (i.e. level, flow, pH, stream temperature, refrigerated

 

 

compartment sample temperature, ORP, rainfall, and any externally logged data - up to 7

 

 

external channels). Also, up to 200 events are logged, including alarm conditions, program

 

 

run/stop events, etc.

 

 

 

 

 

When equipped with integral flow meter, pH/temperature/ORP meter, conductivity, and/or

 

Set Point Sample Trigger

D.O. monitoring options, sampling can be triggered upon an upset condition when field

 

 

selectable limits are exceeded. Concurrent with normal sampling routine, sample liquid is

 

 

deposited in designated “trouble bottle(s)”.

 

 

 

 

 

Multiple Bottle Time, Multiple Bottle Flow, Composite Multiple Bottle Time, Composite

 

Sampling Modes

Multiple Bottle Flow, Composite Time, Composite Flow, Flow with Time Override, Variable

 

 

Interval, Start/Stop, and Level Actuation.

 

 

 

 

Program Delay

Three formats: (1) 1-9,999 minutes or flow pulses in one unit increments (2) programmable

 

sampler start time/date, and (3) programmable time/day of week.

 

 

 

 

 

 

Overload Protection

Controller: 5 amp dc line fuse. Compressor: Thermal overload relay opens at

 

110 °C (230 °F).

 

 

 

 

 

 

Diagnostics

Tests keypad, display, ROM, pump, liquid sensors, velocity signal, and distributor.

 

 

 

 

Program Lock

Access code protection precludes tampering.

 

 

 

Communications

 

 

 

 

 

EPROM Flash Memory

Via RS232. Permits embedded software upgrades in the field.

 

 

 

 

Serial Interface

RS232 compatible; allows on-site collection of stored data.

 

 

 

 

Modem (optional)

14400, V.32 bis, V.42, MNP2-4 error correction. V.42 bis MNP5 data compression.

 

MNP10-EC Cellular Protocol.

 

 

 

 

 

 

Pager

Alarm codes sent to up to three separate pagers’ telephone numbers.

 

 

 

 

 

 

 

 

Page 9

8990spc.fm

Specifications

Page 11
Image 11
Hach 900 MAX manual Specifications